The device was not returned to olympus medical systems corp.(omsc), but returned to the department of olympus (b)(4) for the evaluation.The exact cause of the reported event could not be conclusively determined at this time.If significant additional information is received, this report will be supplemented.
|
Based on the following, olympus re-evaluated the events that occurred on this device and determined that these failure modes were not mdr reportable malfunctions.Therefore, this report is being submitted to retract the mdr on the event.Further inspection of the device by olympus italia s.R.L.Confirmed followings: the cable support in the bending section was disconnected.The bending section was damaged.The first check blocked the movement of the bending section, but it was released by epeating the maneuver.Based on these results, olympus medical systems corp.(omsc) determined that it was possible to unlock the bending section, although it did not bend / return smoothly.
|