SMITH & NEPHEW, INC. LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
|
Back to Search Results |
|
Model Number 72200568 |
Device Problems
Failure to Power Up (1476); No Apparent Adverse Event (3189)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference case-(b)(4).
|
|
Event Description
|
It was reported that the light source, 500 xl xenon had a power supply problem.No case involved.Therefore, there was no patient involved.Results of investigation have concluded that this unit was not powering up which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H3,h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed and no deficiencies were observed.A functional evaluation revealed that the unit would not power up.The line fuses tested as good.The complaint has been confirmed and the root cause has been associated with component failure.Factors unrelated to the manufacturing and design of the device which could have contributed to the reported event, include a defective power supply/ballast or other related electrical components.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
|
|