• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-95
Device Problem Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-95 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error message during procedure.The side affected is unknown.A stirrer motor malfunction can't be excluded.Reportedly, the procedure was completed by a back up device.There was no report of patient injury.
 
Manufacturer Narrative
H.10: a livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue and found out that the stirrer motor was blocked.The stirrer motor was replaced and the problem solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The most likely root cause of the reported event was the corrosion of the motor bearing due to condensation or high temperatures and high level of humidity in the stationary phase which led to the motor shaft blockage.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key11065227
MDR Text Key223434568
Report Number9611109-2020-00714
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-95
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/03/2020
Initial Date FDA Received12/23/2020
Supplement Dates Manufacturer Received02/01/2021
Supplement Dates FDA Received03/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-