• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ERC TUBING CLAMP / 620MM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ERC TUBING CLAMP / 620MM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-05-65
Device Problem Difficult to Open or Close (2921)
Patient Problem No Patient Involvement (2645)
Event Date 12/07/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 erc tubing clamp / 620mm.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 erc tubing clamp / 620mm was giving an error code associated to the clamp jaw position during priming.The issue could be solved by device restarting and was used for the procedure with no further issues.There was no patient involvement.
 
Manufacturer Narrative
H.10: a livanova field service representative was dispatched to the facility to investigate the device and could not confirm the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The most likely root cause of the reported error is a temporary loose connection or by the user extracting the tube.No further complaints have been received for this device/issue.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ERC TUBING CLAMP / 620MM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key11065284
MDR Text Key223759778
Report Number9611109-2020-00716
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-05-65
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/07/2020
Initial Date FDA Received12/23/2020
Supplement Dates Manufacturer Received01/31/2021
Supplement Dates FDA Received02/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-