• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRI-SIL I.C. ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRI-SIL I.C. ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 1758SI18
Device Problem Failure to Infuse (2340)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the catheter balloon was difficult to inflate during pretest.
 
Manufacturer Narrative
The reported event was unconfirmed because the reported failure could not be reproduced.The device met relevant specifications.The product was not used for treatment purposes.The product had not caused the reported failure.Visual evaluation of the returned sample noted one opened (with original packaging), used silicone foley catheter.Visual inspection of the sample noted no obvious visible defects.No root cause could be found because the reported event was unconfirmed.As the event was unconfirmed, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "[warnings] 1.Method for use (1) do not inflate the balloon in the urethra.[the urethra may be injured.] (2) do not pull the catheter hard.[the bladder/urethra may be injured.] 2.Applicable patients ·patients with delirium who might pull out catheter [when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.] [contraindications] 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) this device contains 10% povidone-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients with known allergy to silver coated catheter]" h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that the catheter balloon was difficult to inflate during pretest.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX LUBRI-SIL I.C. ALL-SILICONE FOLEY CATHETER
Type of Device
SILICONE FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key11066078
MDR Text Key223423164
Report Number1018233-2020-21983
Device Sequence Number1
Product Code EZL
Combination Product (y/n)N
PMA/PMN Number
K040504
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 04/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/25/2024
Device Catalogue Number1758SI18
Device Lot NumberNGDQ3190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2020
Initial Date Manufacturer Received 12/02/2020
Initial Date FDA Received12/23/2020
Supplement Dates Manufacturer Received04/11/2021
Supplement Dates FDA Received05/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-