ZIMMER MANUFACTURING B.V. UNKNOWN LINER; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Positioning Failure (1158)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/30/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: foreign country: (b)(6).The device will not be returned for analysis, due to location of device is unknown; however, an investigation of the reported event is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0002648920-2020-00536.
 
Event Description
It was reported that during an initial surgery, the c-ring in bipolar shell was locked with shell and it would not assemble with polyethylene liner, which led to a 20 minute delay to change into new implant.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: h2; h6 the liner was not returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were not provided.Part and lot identification of the liner are necessary for review of device history records, neither were provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN LINER
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
MDR Report Key11067170
MDR Text Key224323409
Report Number0002648920-2020-00537
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 05/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/30/2020
Initial Date FDA Received12/23/2020
Supplement Dates Manufacturer Received12/30/2020
05/04/2021
Supplement Dates FDA Received01/20/2021
05/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
D11: ZIMMER SHELL CAT#00500104700 LOT#62779209
Patient Age75 YR
-
-