Catalog Number 8065977763 |
Device Problems
Crack (1135); Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/24/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.(b)(4).
|
|
Event Description
|
A nurse reported that during an intraocular lens (iol) implant procedure, a cartridge cracked as the lens was inserted and resulted in a fractured tip.There was no impact on the lens.There was patient contact but no patient impact.The surgery was completed the same day.
|
|
Event Description
|
Additional information was provided that the surgeon only uses certain lens models.These lens models are only qualified for use with other company cartridges.It is unknown if a qualified handpiece was used.
|
|
Manufacturer Narrative
|
Six company cartridge cartons were returned for the reported company cartridge lot# 15096777 related to three complaint files.The used complaint company cartridge was not returned.Initially, two opened and four unopened company cartridge cartons were returned for the company cartridge lot # 15096777.There were a total of seventy-nine samples returned.Eight samples were randomly pulled from the seventy-nine unopened company cartridge samples returned for lot# 15096777.The eight company cartridges were microscopically examined and no damage or abnormalities were observed.The company cartridges were functionally tested per the dfu.No lens or cartridge damage was observed after the lens delivery.The cartridge was cleaned for further evaluation.Top coat due stain testing was conducted with acceptable results.The company cartridge product history records were reviewed and the documentation indicated the product met release criteria.Information was provided that the surgeon only uses certain lens models.These lens models are only qualified for use with other company cartridges.It is unknown if a qualified handpiece was used.The root cause for the reported may be related to a failure to follow the dfu.New information was provided that the surgeon only uses certain lens models.These models are only qualified for use with other company cartridges.The use of non-qualified combinations may lead to delivery issues and/or damage to the lens or the cartridge.The used complaint company cartridges were not returned for evaluation.Eight unopened samples for the reported lot number were evaluated.No damage or abnormalities were observed.The eight samples were functionally evaluated per the dfu.No lens or cartridge damage was observed.Top coat due stain testing was conducted with acceptable results.It is unknown if a qualified handpiece was used.The manufacturer internal reference number is: (b)(4).
|
|
Manufacturer Narrative
|
Six company cartridge cartons which contained a total of seventy-nine unopened samples were returned for the reported company cartridge lot# 15096777.The used complaint company cartridge was not returned.Two opened and four unopened company cartridge cartons were returned for the company cartridge lot # 15096777.There were a total of seventy-nine unopened samples returned.Eight samples were randomly pulled from the seventy-nine unopened company cartridge samples returned for lot# 15096777.The eight company cartridges were microscopically examined and no damage or abnormalities were observed.The company cartridges were functionally tested per the dfu.No lens or cartridge damage was observed after the lens delivery.The cartridge was cleaned for further evaluation.Top coat due stain testing was conducted with acceptable results.The company cartridge product history records were reviewed and the documentation indicated the product met release criteria.It is unknown if a qualified lens model/diopter and handpiece were used.The root cause for the reported damage cannot be determined.The used complaint company cartridge was not returned for evaluation.Eight unopened samples for the reported lot number were evaluated.No damage or abnormalities were observed.The eight samples were functionally evaluated per the dfu.No lens or cartridge damage was observed.Top coat due stain testing was conducted with acceptable results.It is unknown if a qualified lens model/diopter and handpiece were used.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|