Catalog Number 8065751761 |
Device Problem
Insufficient Information (3190)
|
Patient Problem
Burn, Thermal (2530)
|
Event Date 12/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The sample has been received and in-house evaluation is in progress.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
|
|
Event Description
|
A health professional reported the surgeons hand was burned from the phacoemulsification handpiece.Procedure details or if surgeon required any treatment for the burn have not been provided.Additional information has been requested but not received.
|
|
Manufacturer Narrative
|
The manufacturer internal reference number is: (b)(4).
|
|
Manufacturer Narrative
|
The phacoemulsification (phaco) handpiece was received for evaluation.The returned phaco handpiece was connected to a calibrated system.A system message (sm) [tune failed ¿ loose tip] displayed when the phaco handpiece attempted tuning, and further testing could not be performed.Disassembling the phaco handpiece revealed moisture ingress within the electrode chamber.Unrelated to the reported event, moisture ingress was found to cause a short circuit from the high electrode to ground; however, how or when moisture entered the phaco handpiece remains inconclusive.The phaco handpiece manufacturing device history record (dhr) was reviewed.Based on assessment, the product met specifications at the time of release.The root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
|
|