Catalog Number 8065751761
Device Problem Insufficient Information (3190)
Patient Problem Burn, Thermal (2530)
Event Date 12/01/2020
Event Type  malfunction  
Manufacturer Narrative
The sample has been received and in-house evaluation is in progress.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A health professional reported the surgeons hand was burned from the phacoemulsification handpiece.Procedure details or if surgeon required any treatment for the burn have not been provided.Additional information has been requested but not received.
 
Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The phacoemulsification (phaco) handpiece was received for evaluation.The returned phaco handpiece was connected to a calibrated system.A system message (sm) [tune failed ¿ loose tip] displayed when the phaco handpiece attempted tuning, and further testing could not be performed.Disassembling the phaco handpiece revealed moisture ingress within the electrode chamber.Unrelated to the reported event, moisture ingress was found to cause a short circuit from the high electrode to ground; however, how or when moisture entered the phaco handpiece remains inconclusive.The phaco handpiece manufacturing device history record (dhr) was reviewed.Based on assessment, the product met specifications at the time of release.The root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key11071932
MDR Text Key224341723
Report Number2028159-2020-01154
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 05/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751761
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2020
Initial Date Manufacturer Received 12/07/2020
Initial Date FDA Received12/24/2020
Supplement Dates Manufacturer Received03/29/2021
04/22/2021
Supplement Dates FDA Received03/29/2021
05/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CENTURION VISION SYSTEM
-
-