Device identifier: (b)(4).The root cause cannot be determined at this time.Results of the documentary investigation does not reveal any issue about the design, the manufacturing, during the use.No trend for this kind of incident is observed.Associated risk is correctly evaluated.Product is currently not returned to integra so no failure analysis is possible at this time.Following previous similar incidents, it was observed that the required loosening torque during hallu lock removal is more important cause of the bone growth around the implants and can reach the mechanical limit of the raw material.Furthermore, due to the size of the associated screws, the dimensions of the tip of screwdriver 219127nd are very tiny and per consequence fragile when it used repeatedly and / or with important torque.For removal procedure, a dedicated removal kit is available with 2 screwdrivers p/n 219127nd that are checked between each removal surgery in order to limit the observed risks.
|