SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 102096-67A
Device Problem Imprecision (1307)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/25/2020
Event Type  malfunction  
Manufacturer Narrative
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
 
Event Description
On novemeber 26th 2020,senseonics was made aware of an incident where user experienced inaccuacies in sensor readings which leads to early sensor removal.
 
Manufacturer Narrative
Per initial investigation analysis, it was observed that there was change in the system diagnostics, due to which the system asserted multiple sensor check alerts, after the data was synced.Based on the data it could not be confirmed if the system will recover.The user had stopped using the system per the case notes, so an rma was authorized due to customer experience.Since the rma has not been received, there could not be any further investigations done at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20875 7005
MDR Report Key11073651
MDR Text Key228276265
Report Number3009862700-2020-00597
Device Sequence Number1
Product Code QCD
UDI-Device Identifier00817491022349
UDI-Public00817491022349
Combination Product (y/n)Y
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 11/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/05/2020
Device Model Number102096-67A
Device Catalogue NumberFG-4200-00-301
Device Lot NumberWP07609
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/26/2020
Initial Date FDA Received12/26/2020
Supplement Dates Manufacturer Received11/26/2020
Supplement Dates FDA Received02/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age50 YR
-
-