• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-Q180V
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/02/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user that during the inspection by the biomedical engineer, the forceps elevator could not move down at all.Olympus (b)(4) checked the subject device and found that the reported phenomenon could not duplicated.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation, but was returned to olympus australia (oaz).Oaz inspected the device and the reported phenomenon was not duplicated.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the information from oaz there was the possibility that the reported phenomenon was attributed to the following occurrence mechanism.- the movement of the forceps elevator was temporarily hindered due to foreign object adhering to the area around the forceps elevator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11073996
MDR Text Key223800537
Report Number8010047-2020-11025
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170339967
UDI-Public04953170339967
Combination Product (y/n)N
PMA/PMN Number
K143153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-Q180V
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/02/2020
Initial Date FDA Received12/27/2020
Supplement Dates Manufacturer Received02/08/2021
Supplement Dates FDA Received02/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-