Model Number OER-PRO
Device Problem Decrease in Pressure (1490)
Patient Problem No Patient Involvement (2645)
Event Date 12/02/2020
Event Type  malfunction  
Manufacturer Narrative
An olympus field service engineer (fse) was dispatched to confirm the reported issue.The fse found that the incoming water supply was lower than normal.The fse found the internal water filter clogged.The fse observed the filter replacement, water line disinfection, chemical replacement and heat up and one full cycle.The cycle time was back to the expected 27-28 minutes.The device was repaired and verified according to specifications.Software attributes were verified and confirmed.If additional information is obtained, a supplemental report will be filed.
 
Event Description
A user facility reported that the timer on an endoscope reprocessor unit was incorrect.There was no patient involvement or injuries reported.No additional information has been obtained.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The following sections were updated: g4, g7, h2, h4, h6 and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the water filter clogged because it was used for more than the recommended period without replacement.The following is stated in the instructions for use and the event can be prevented: "chapter 7.2.Replacing the water filter (maj-824), replace the water filter at least once a month to prevent contamination of the rinse water.The water filter should also be replaced whenever an error code indicating water supply insufficiency [e01] is displayed.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11074012
MDR Text Key223798204
Report Number8010047-2020-11029
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/02/2020
Initial Date FDA Received12/27/2020
Supplement Dates Manufacturer Received02/17/2021
Supplement Dates FDA Received02/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-