• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CURRENT HEALTH LTD CURRENT HEALTH LTD SECOND GENERATION; TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CURRENT HEALTH LTD CURRENT HEALTH LTD SECOND GENERATION; TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY Back to Search Results
Catalog Number CH0535C
Device Problem Overheating of Device (1437)
Patient Problem Partial thickness (Second Degree) Burn (2694)
Event Date 12/14/2020
Event Type  Injury  
Event Description
Upon home health nurse visit on (b)(6) 2020 patient reported she was burned by the unit sensor on her upper outer left arm.Second degree burn noted, no medical treatment required by md or emergency department.Notified current health contact and was instructed to ship the unit back to the company for testing and evaluation.Shipped back to company on (b)(6) 2020 according to manufacture requirements.Reported up to home care/hospital risk management, senior leadership, patient physician, and all appropriate reports completed.All additional generation 2 current health telehealth systems removed from home care patients until evaluation and testing is completed by manufacturer.Manufacturer in full support of removing all units until testing completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURRENT HEALTH LTD SECOND GENERATION
Type of Device
TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY
Manufacturer (Section D)
CURRENT HEALTH LTD
edinburgh
UK 
MDR Report Key11076514
MDR Text Key224100577
Report NumberMW5098542
Device Sequence Number1
Product Code DRG
Combination Product (y/n)Y
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberCH0535C
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/24/2020
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age77 YR
Patient Weight60
-
-