This report is being supplemented to provide additional information based on the legal manufacturer's investigation and to correct date included on the initial report.A review of the device history record confirmed the subject device was shipped in accordance to specifications.Based on the results of the legal manufacturer's investigation, the phenomenon likely occurred due to the lid coming in contact with a scope or hard object.The phenomenon could be detected by handling the device in accordance to the verbiage identified in the instructions for use below: "chapter 3 inspection before use - 3/2 inspecting the lid and the lid packing: before using the equipment, always check that there is no irregularity regarding the following points on the lid and the lid packing.If there is any irregularity, cleaning fluid or disinfectant solution may leak out.".
|