• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number JC8537
Device Problem Disconnection (1171)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Event date: the event occurred on (b)(6) 2020.(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing of a clearlink system continu-flo solution set disconnected from the valve.This issue was identified during setup and preparation prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to d10, h3 and h6.H10: the device was received for evaluation.Visual inspection was performed using the naked eye and the tubing was found separated from the second y-site in the upstream part.Additional inspection was performed which observed that the solvent was not uniformly applied in the circumference of the tubing.Dimensional testing was also performed, and it was determined that the components were as according to specification.The reported condition was verified.The cause of the reported condition was due to lack of solvent or an inadequate insertion during assembly of the second y site and the tubing.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11082857
MDR Text Key225167988
Report Number1416980-2020-08081
Device Sequence Number1
Product Code FPA
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberJC8537
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2021
Initial Date Manufacturer Received 12/03/2020
Initial Date FDA Received12/28/2020
Supplement Dates Manufacturer Received01/29/2021
Supplement Dates FDA Received02/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-