• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 21AGFN-756
Device Problem Material Separation (1562)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2020
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.The investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2020, a sjm regent heart valve w/ flex cuff procedure took place.When the physician went to rotate the valve, a leaflet broken off.The piece was recovered and the patient remained stable.
 
Manufacturer Narrative
The reported event of a dislodged leaflet was confirmed.Both leaflets were dislodged from the orifice intact and returned with the valve.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the leaflet dislodgment could not be conclusively determined.Please note, per the instructions for use artmt100122074 ver.A, "using the valve holder/rotator and the flexible valve holder handle model 905-hh, or the rigid valve holder handle model 905-rhh, rotate the valve in situ to the desired position.The valve should rotate freely.If resistance is noted, the valve holder/rotator may not be properly seated in the valve, the valve may not be in the fully closed position, or the valve may be oversized.If the valve does not rotate freely, do not force valve rotation.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE W/FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key11083615
MDR Text Key227355313
Report Number2648612-2020-00140
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005975
UDI-Public05414734005975
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21AGFN-756
Device Catalogue Number21AGFN-756
Device Lot Number7625223
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/04/2020
Initial Date FDA Received12/28/2020
Supplement Dates Manufacturer Received01/21/2021
Supplement Dates FDA Received01/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-