MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number JC8537
Device Problem Disconnection (1171)
Patient Problem No Patient Involvement (2645)
Event Date 12/02/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing of a clearlink system continu-flo solution set disconnected from the valve.This issue was identified during setup and preparation prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the device was received, for evaluation without the tubing part.Visual inspection was performed using the naked eye, which observed the tubing was separated from the first y-site clearlink component.Additional inspection was performed, which observed a lack of solvent in the clearlink housing.Dimensional testing was performed, at the clearlink housing.And the component was according to specification.The reported condition was verified.The cause of the condition could not be determined, as the tubing was not available for further evaluation.However, the most likely cause, was due to the manufacturing process.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11084274
MDR Text Key225286316
Report Number1416980-2020-08112
Device Sequence Number1
Product Code FPA
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberJC8537
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2021
Initial Date Manufacturer Received 12/03/2020
Initial Date FDA Received12/28/2020
Supplement Dates Manufacturer Received01/29/2021
Supplement Dates FDA Received02/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-