MAUDE Adverse Event Report: V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems Failure to Align (2522); Key or Button Unresponsive/not Working (4063)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 28dec2020.
 
Event Description
The customer reported while attempting to perform a pre-use check prior to being put into service.The screen did not react to touch.The incident happened prior to being put into use.There was no patient involvement or delay in treatment.No patient/user harm was reported.The field service engineer (fse) confirmed misalignment in the touch position reported failure.The issue was not resolved by calibrating the touchscreen; therefore, the touchscreen was replaced.
 
Manufacturer Narrative
G4:20apr2021.B4:26apr2021.The touchscreen was returned to manufacturer for failure investigation analysis.The reported complaint touchscreen failed is not duplicated.There is no fault found on this returned touchscreen assembly.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key11084650
MDR Text Key225276980
Report Number2031642-2020-04756
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/30/2020
Initial Date FDA Received12/28/2020
Supplement Dates Manufacturer Received11/30/2020
Supplement Dates FDA Received04/26/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/03/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-