• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE RP PS ARTIC SURF SZ8; ATTUNE INSTRUMENTS : TIBIAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE RP PS ARTIC SURF SZ8; ATTUNE INSTRUMENTS : TIBIAL TRIALS Back to Search Results
Model Number 2545-00-548
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The peg w/o the spring has broken off.The peg with the spring has the top edge cracked off.Both pieces were found after washing the set during inspection.Unsure of where broken pieces are.Both of these items must have been this way for sometime as we have not used these sizes in a few weeks.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirmed the reported event.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP PS ARTIC SURF SZ8
Type of Device
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key11089476
MDR Text Key224153353
Report Number1818910-2020-27802
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295132882
UDI-Public10603295132882
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2545-00-548
Device Catalogue Number254500548
Device Lot NumberMVMBSV750
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2020
Initial Date Manufacturer Received 12/14/2020
Initial Date FDA Received12/29/2020
Supplement Dates Manufacturer Received01/07/2021
Supplement Dates FDA Received01/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-