• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES ANSPACH EMAX 2 PLUS HAND PIECE - ROHS; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES ANSPACH EMAX 2 PLUS HAND PIECE - ROHS; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number PFSR101209
Device Problem Interrogation Problem (4017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2020
Event Type  malfunction  
Event Description
It was reported that during a navio saw bone demo, it was found that on connecting the anspach drill to the console, the e6 error is activated (overtemp error that indicates the drill is faulty).No patient was involved.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the navio anspach drill, part number 101209, serial number (b)(6), intended for use in treatment, was returned for evaluation.A relationship between the reported event and the device was confirmed.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported problem was confirmed.Upon connection to a known good navio system, an e6 error was displayed.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most likely cause of this event is component failure.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANSPACH EMAX 2 PLUS HAND PIECE - ROHS
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11090004
MDR Text Key224303579
Report Number3010266064-2020-02138
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPFSR101209
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/06/2020
Initial Date FDA Received12/29/2020
Supplement Dates Manufacturer Received07/23/2021
Supplement Dates FDA Received07/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NAVIO SURGICAL SYSTEM INDIA ROB00036
-
-