Catalog Number 2C2009K |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that 4 large volume infusors were leaking from the end of the port.The leaks were identified during unspecified process steps.It was not specified if a patient was connected for therapy at the time of the event.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the actual device was not available; however, twenty-three (23) companion samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional leak testing was performed on all the samples and no issue were noted.Based on the evaluation findings, the 23 companion samples were determined to be conforming product.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Additional information was added to d4: lot #: and udi#: and h6:.D4: lot #: 20d033, previously submitted as ¿asku¿; d4: udi#: (b)(4), previously submitted as ¿ni¿; h4: the device was manufactured from april 9, 2020 - april 10, 2020.H10: a batch review was conducted, and there were no deviations found related to this reported condition, during the manufacture of this lot.The device was not received for evaluation.Therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the devices were discarded and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|