• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C2009K
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that 4 large volume infusors were leaking from the end of the port.The leaks were identified during unspecified process steps.It was not specified if a patient was connected for therapy at the time of the event.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the actual device was not available; however, twenty-three (23) companion samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional leak testing was performed on all the samples and no issue were noted.Based on the evaluation findings, the 23 companion samples were determined to be conforming product.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information was added to d4: lot #: and udi#: and h6:.D4: lot #: 20d033, previously submitted as ¿asku¿; d4: udi#: (b)(4), previously submitted as ¿ni¿; h4: the device was manufactured from april 9, 2020 - april 10, 2020.H10: a batch review was conducted, and there were no deviations found related to this reported condition, during the manufacture of this lot.The device was not received for evaluation.Therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the devices were discarded and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11090607
MDR Text Key224757603
Report Number1416980-2020-08141
Device Sequence Number1
Product Code MEB
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 03/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C2009K
Device Lot Number20D033
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/03/2020
Initial Date FDA Received12/29/2020
Supplement Dates Manufacturer Received01/14/2021
03/09/2021
03/18/2021
Supplement Dates FDA Received01/28/2021
03/12/2021
03/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-