SMITH & NEPHEW, INC. OR3O DUAL MOBILITY TRIAL LINER 42/54; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 71353304
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported during thr procedure that the trial was broken outside the patient.It is unknown if there was a delay and how was the procedure finished.No patient injury or other complications were reported.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.A piece of the trial is chipped/broken, rendering device inoperable.The device was manufactured in 2019 and shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OR3O DUAL MOBILITY TRIAL LINER 42/54
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11091084
MDR Text Key224244493
Report Number1020279-2020-07947
Device Sequence Number1
Product Code LPH
UDI-Device Identifier00885556710470
UDI-Public885556710470
Combination Product (y/n)N
PMA/PMN Number
K191002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71353304
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2020
Initial Date Manufacturer Received 12/07/2020
Initial Date FDA Received12/29/2020
Supplement Dates Manufacturer Received02/01/2021
Supplement Dates FDA Received02/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-