• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Device Problem Migration (4003)
Patient Problem Failure of Implant (1924)
Event Type  malfunction  
Manufacturer Narrative
Even though no product has been returned due to the device remaining implanted, an fsca investigation was performed and variation in the torque applied to the threaded cap during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although all operators followed the assembly procedure, and the torque wrench indicated 40 in-lbs was applied, the manner of using the torque wrench resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap, thereby creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.
 
Event Description
Information was received that x-ray images revealed that the end cap had fallen off.No revision procedure has been performed at this time.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
Manufacturer Contact
tiara rae
101 enterprise dr, suite 100
aliso viejo, CA 92656
6197314188
MDR Report Key11092299
MDR Text Key224388225
Report Number3006179046-2020-00525
Device Sequence Number1
Product Code PGN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/09/2020
Initial Date FDA Received12/29/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1898-2020
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age5 YR
Patient Weight19
-
-