Model Number URF-P7 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/25/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device has not been returned to omsc but was returned to olympus (b)(4).(b)(4) sent the device to a third party laboratory for microbiological testing.As a result of the testing, no microbe was detected from the sample collected from the all channels of the subject device.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that as a result of routine microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device.Unspecified microbes (11 cfu/endoscope).Other detailed information such as the reprocessing method was not provided.There was no report of infection associated with this report.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.Olympus contacted with the user and got the information that the device had been manually reprocessed with peracetic acid (paa).Omsc reviewed the manufacture history (dhr) of the device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.
|
|