Catalog Number 71674076 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that during use, inside the patient, the t-handle broke.The procedure finished with a competitor device.Surgical delay less than 30 minutes.Patient injuries were not reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The device, used in treatment was returned for evaluation.Visual inspection of the returned device confirms that the device has signs of wear and tear from use.The device was manufactured in 2011.According to clinical/medical investigation, although it was reported the t-handle broke during use internal to the patient, per subsequent e-mail it was reported nothing was left inside of the patient.According to the report, the procedure finished with a competitor device with a surgical delay of less than 30 minutes.Since there were no patient injuries reported, no further clinical/medical assessment is warranted at this time.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|