Model Number CD2357-40Q
Device Problems Incorrect Measurement (1383); Compatibility Problem (2960)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/21/2020
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the an asymptomatic patient presented to the clinic for an mri on (b)(6) 2020.The patient's implantable cardioverter defibrillator was placed into mri mode, which caused the estimated battery longevity to be much lower than expected.A follow-up remote transmission showed that the estimated battery longevity was back to normal readings.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key11095791
MDR Text Key224390220
Report Number2017865-2020-25136
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734508094
UDI-Public05414734508094
Combination Product (y/n)N
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Model NumberCD2357-40Q
Device Catalogue NumberCD2357-40Q
Device Lot NumberA000097347
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/21/2020
Initial Date FDA Received12/30/2020
Supplement Dates Manufacturer Received12/21/2020
Supplement Dates FDA Received01/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age35 YR
-
-