MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION 2000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION 2000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938740
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 12/05/2020
Event Type  malfunction  
Manufacturer Narrative
Initial reporter facility name: (b)(6).Initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 2000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag leaked.This was further described as one unit ¿had external fluid leakage from the tube¿.This issue was detected during setup and preparation prior to patient use.There was no patient involvement.As a result, the product was replaced and a new product was used to continue the treatment.No additional information is available.
 
Manufacturer Narrative
H4: device manufactured between january 31, 2020 to february 01, 2020.H10:the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed which revealed a leak at the spike port bonding area only.The reported condition was verified.The cause was not determined; however the most likely cause was due to inadequate or lack of cyclohexanone being applied to the cap tubing when it was inserted to the spike port during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11096219
MDR Text Key224537280
Report Number1416980-2020-08202
Device Sequence Number1
Product Code LHI
Combination Product (y/n)Y
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938740
Device Lot Number60225383
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2020
Initial Date Manufacturer Received 12/08/2020
Initial Date FDA Received12/30/2020
Supplement Dates Manufacturer Received01/19/2021
Supplement Dates FDA Received02/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-