Additional information was added to d10, h3, h4 and h6.H4: the lot was manufactured from june 23, 2020 - june 24, 2020.H10: the device was received for evaluation.The upper left corner was cut where the customer likely drained the contents.Unaided visual inspection was performed which observed a tear at the middle area of the left side seam of the bag.Functional testing was performed with water which revealed a leak from the affected area approximately at the 800 ml volume level.The reported condition was verified.The cause of the tear/hole could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|