Generator analysis was completed and the high impedance was not replicated on the generator as various electrical loads were attached to the generator and diagnostic tests were normal.An automated evaluation showed that the device functioned as expected.Lead analysis was completed and braded openings were noted in the outer silicone tubing.The reported of lead fracture were verified.A break was identified in the negative coil.Scanning electromagnetic images of the negative coil show putting or electro-etching conditions had occurred in the break location.The coil appearance also suggested a stress-induced fracture in at least one strand of the quadfilar coil.Note that since the furthest electrode to the bifurcation was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Other than the above mentioned observations and typical wear and explant related observations, no other anomalies were identified in the returned lead portions.No additional relevant information has been received to date.
|