• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020, date of report: 30dec2020.
 
Event Description
The field service engineer (fse) found the power cord was out of spec during preventive maintenance (pm).The field service engineer (fse) was able to duplicate the reported problem.The field service engineer reported that the unit was not in use on patient.The issue was found during performance verification test (pvt).The field service engineer replaced the power cord to address the reported problem.
 
Manufacturer Narrative
G4: (b)(6) 2021.B4: (b)(6) 2021.Failure analysis on the returned ac power cord shows that the ac power cord failed esa resistance test.Verified reported failure.High pin-pin resistances noted n-n and l-l.Root cause determined to be mechanical damage.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key11096742
MDR Text Key225337087
Report Number2031642-2020-04803
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 12/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/02/2020
Initial Date FDA Received12/30/2020
Supplement Dates Manufacturer Received12/02/2020
Supplement Dates FDA Received04/26/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/18/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-