Model Number 66800041 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that, the versajet exact assy, 45 degree x 14mm was leaking and the fluid jet was inconsistent.It is unknown when the event occur or if a patient was involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The device, intended for use in treatment, was returned for evaluation.A visual inspection found there not to be an obstruction at the distal output orifice.The functional evaluation revealed there was no flow when the feed line was connected to the fluid supply.The device did not prime when the test console was ramped up, establishing a relationship between the device and the reported event.The root cause was identified as a component failure.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found other related events.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
|
|