Model Number V60
Device Problem Unexpected Shutdown (4019)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020, date of report: 30dec2020.
 
Event Description
The customer reported the ventilator shut down while in use.The device was ventilating a patient at the time of the shutdown.
 
Manufacturer Narrative
G4:25mar2021.B4:02apr2021.The customer clarified that the ventilator was being set up for patient use when the ventilator shut down.The ventilator was not yet connected to the patient.Respiratory therapy staff swapped the ventilator out for a different v60 ventilator.There was no patient harm or delay to therapy.The philips field service engineer (fse) evaluated the unit and determined that the power management board needed to be replaced.The fse replaced the power management board and the issue was resolved.The ventilator successfully passed performance verification testing after the repair was completed.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key11097058
MDR Text Key225332026
Report Number2031642-2020-04808
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Remedial Action Recall
Type of Report Initial,Followup
Report Date 12/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/03/2020
Initial Date FDA Received12/30/2020
Supplement Dates Manufacturer Received12/03/2020
Supplement Dates FDA Received04/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/26/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1622-2020
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER
-
-