• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD AIRVO 2 HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD AIRVO 2 HUMIDIFIER Back to Search Results
Model Number PT101
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint airvo 2 humidifier was received at our fisher & paykel healthcare (f&p) regional office in the (b)(6) and was inspected by a trained f&p technician.The device was performance tested and the audible alarm function was checked.Results: during testing it was found that the audible alarm did not function.Previous investigations into this type of failure have identified that the problem is caused by a faulty speaker and electrical resistance testing has shown the speaker's resistance to be open circuit.Conclusion: as part of our ongoing product improvement initiatives, a new speaker unit has been sourced from a different supplier.The airvo 2 humidifier user manual states that the "airvo 2 is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases" and that "the unit is not intended for life support." the user manual warns the user: - prior to each patient use, ensure that the auditory alarm signal is audible by conducting the alarm system functionality check described in the alarms section the alarm system functionality check instructs the user on how to test the alarm and states that "if either alarm signal is absent, do not use the unit.Contact your fisher & paykel healthcare representative.".
 
Event Description
A healthcare facility in (b)(6) reported an issue with a pt101 airvo 2 humidifier.Upon device assessment at our fisher & paykel healthcare (f&p) regional office in the (b)(6), it was found that the audio alarm of the pt101 airvo 2 humidifier was not functioning.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRVO 2 HUMIDIFIER
Type of Device
AIRVO 2 HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
MDR Report Key11097683
MDR Text Key224529922
Report Number9611451-2020-01221
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
PMA/PMN Number
K131895
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 12/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPT101
Device Catalogue NumberPT101
Device Lot Number1509220201
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/03/2020
Initial Date FDA Received12/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/22/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-