• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C1009K
Device Problem No Flow (2991)
Patient Problem No Patient Involvement (2645)
Event Date 12/08/2020
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone no.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a large volume infusor did not flow during filling.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4 and h6.H4: the device was manufactured from december 20, 2019 - december 21, 2019.H10:the actual device was not available.However, a photograph of the sample was provided for evaluation.Functional flow rate testing must be performed, on the actual device in order to verify, the accuracy of the alleged device.Therefore, the reported event could not be objectively verified.A batch review was conducted, and there were no deviations found related to this reported condition, during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: visual inspection of the photograph did not identify any abnormalities that could have contributed to the reported condition.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11097794
MDR Text Key224535579
Report Number1416980-2020-08225
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412022529
UDI-Public(01)00085412022529
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 04/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C1009K
Device Lot Number19N037
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/08/2020
Initial Date FDA Received12/30/2020
Supplement Dates Manufacturer Received01/14/2021
03/10/2021
Supplement Dates FDA Received02/04/2021
04/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-