Additional information was added to h3, h4 and h6.H4: the device was manufactured from april 9, 2020 - april 10, 2020.H10: the device was received for evaluation.A visual inspection was performed, and it was noted that fluid was inside the bag that contain the unit which suggested a leak had occurred.A functional leak test was performed, and it was noted that there was evidence of a leak observed coming out at the solvent-bonded junction of the tubing and stressmember.The reported condition was verified.The cause of the reported condition was a manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|