• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS; INTRAOCULAR LENS Back to Search Results
Model Number VICM5_12.6
Device Problem Misfocusing (1401)
Patient Problem Blurred Vision (2137)
Event Date 06/24/2020
Event Type  Injury  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).No similar complaint type events reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 12.6mm vicm5_12.6 implantable collamer lens, -10.0 diopter into the patient's left eye (os) on (b)(6) 2020.On (b)(6) 2020 the lens was exchanged for a same size, different model lens due to refractive surprise.The problem was resolved.The cause of the event is reported as a patient-related factor.
 
Manufacturer Narrative
Additional information: h3- device evaluation: lens was returned in a micro centrifuge vial with moisture and foreign residue/debris.Visual inspection found no visible damage.Dimensional and functional inspection found the lens to be within specifications.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key11098617
MDR Text Key224513439
Report Number2023826-2020-03256
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model NumberVICM5_12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2020
Initial Date Manufacturer Received 12/08/2020
Initial Date FDA Received12/30/2020
Supplement Dates Manufacturer Received04/23/2021
Supplement Dates FDA Received05/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CARTRIDGE MODEL-SFC-45, LOT#-UNK; FOAM TIP PLUNGER MODEL-FTP, LOT#-UNK; INJECTOR MODEL-MSI-PF, LOT#-UNK
Patient Outcome(s) Required Intervention;
Patient Age36 YR
-
-