It was reported that the patient presented to the emergency room.Interrogation of the pacemaker was attempted but was unsuccessful using inductive telemetry.The device did not have any pacing response when a magnet was placed over the device.A telemetry test was attempted but was also unsuccessful.It was believed that the device may be dead, and there is no allegation of premature battery depletion from the physician.The device was explanted and replaced, and the patient was in stable condition throughout with no consequences.
|
The reported event of inability to interrogate was confirmed.As received, the device had no telemetry communication and was in backup mode.Visual inspection of the header attachment area detected a bonding anomaly.A device hermeticity breach was observed, consistent with feedthrough damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to the internal electronics.The device was cut open to enable further testing and the battery was found depleted to end-of-service level.Hybrid circuitry was tested indicating elevated current drain, consistent with moisture damage, depleting the battery and resulting in the reported event.A manufacturing anomaly may have occurred, which resulted in the header bonding anomaly.As a result of this finding, abbott is performing further investigation.
|