Model Number PT100
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint pt100 myairvo humidifier was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the information provided by the customer, previous investigations, and our knowledge of the product.Results: the customer has stated that the myairvo unit did not have an audible alarm.Previous investigations into audio alarm failures have identified that the problem is caused by a faulty speaker and electrical resistance testing has shown the speaker's resistance to be open circuit.Conclusion: as part of our ongoing product improvement initiatives, a new speaker unit has been sourced from a different supplier.The myairvo user manual states that the "myairvo is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases" and that "the unit is not intended for life support." the user manual warns the user: - prior to each patient use, ensure that the auditory alarm signal is audible by conducting the alarm system functionality check described in the alarms section.The alarm system functionality check instructs the user on how to check the alarm and states that "if either alarm signal is absent, do not use the unit.Contact your fisher & paykel healthcare representative.".
 
Event Description
A distributor in (b)(4) reported via a fisher and paykel healthcare (f&p) field representative that the audible alarm of a pt100 myairvo humidifier was not functioning.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYAIRVO HUMIDIFIER
Type of Device
MYAIRVO HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
MDR Report Key11099461
MDR Text Key224529843
Report Number9611451-2020-01241
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
PMA/PMN Number
K131895
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial
Report Date 12/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberPT100
Device Catalogue NumberPT100
Device Lot Number2100174739
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/10/2020
Initial Date FDA Received12/30/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-