Model Number N/A |
Device Problems
Difficult to Insert (1316); Mechanical Problem (1384); Material Integrity Problem (2978)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source - (b)(6).Customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that an articular surface has damage and an edge that does not allow proper implantation.Another articular surface was used.There is no additional information available at this time.
|
|
Manufacturer Narrative
|
No product was returned or pictures provided; therefore, visual and dimensional evaluations could not be performed.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.H6: type of investigation - 4114 - device not returned code is n/a which was reported on initial report.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned deivce shows that the dovetail feature is compressed and flared out.Also the proximal and distal surface exhibit damage in various areas of device surface.The root cause is attributed to use error.There are warnings in the package insert that this type of event can occur and risks are addressed in risk documentation.The contact was sent the investigation results.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Search Alerts/Recalls
|