• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) AESPIRE 7100; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) AESPIRE 7100; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2020
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.The distributor performed a checkout of the equipment and confirmed the reported complaint.The cable/harness connection was reseated to resolve the issue.
 
Event Description
The hospital reported an error that could result in an inability to switch ventilation modes as expected.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi 21402 8
CH  214028
Manufacturer Contact
john szalinski
MDR Report Key11100580
MDR Text Key225766364
Report Number9710602-2020-00422
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial
Report Date 12/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/04/2020
Initial Date FDA Received12/31/2020
Date Device Manufactured01/20/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-