ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR
|
Back to Search Results |
|
Model Number AU00T0 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.(b)(4).
|
|
Event Description
|
An intraocular lens (iol) reply card was received from a customer stating that during an iol implant procedure, the lens was defective and not implanted.There was no reported patient harm.Additional information has been requested and received that the doctor stated that the lens was not unfolding correctly.
|
|
Manufacturer Narrative
|
The product was not returned.All product and batch history records are quality reviewed prior to product release.Viscoelastic was not provided.It is unknown if a qualified product was used.The product investigation could not identify a root cause.It is unknown if a qualified viscoelastic was used.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.Information was provided that a facility representative indicated there was currently no other details available, and that information would be provided if anything became available.No further information has been provided.The product was not returned to evaluate.File will be reopened when product is received.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|
|
|