• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the product was not returned for analysis.The product history and batch records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause.There have been no other complaints reported in the lot number.(b)(4).
 
Event Description
An intraocular lens (iol) reply card was received from a customer stating that during an iol implant procedure, the lens was defective and not implanted.There was no reported patient harm.Additional information has been requested and received that the doctor stated that the lens was not unfolding correctly.
 
Manufacturer Narrative
The product was not returned.All product and batch history records are quality reviewed prior to product release.Viscoelastic was not provided.It is unknown if a qualified product was used.The product investigation could not identify a root cause.It is unknown if a qualified viscoelastic was used.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.Information was provided that a facility representative indicated there was currently no other details available, and that information would be provided if anything became available.No further information has been provided.The product was not returned to evaluate.File will be reopened when product is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key11108804
MDR Text Key226065574
Report Number1119421-2021-00003
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Model NumberAU00T0
Device Lot Number12735678
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/15/2020
Initial Date FDA Received01/04/2021
Supplement Dates Manufacturer Received02/13/2021
Supplement Dates FDA Received02/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-