ZIMMER BIOMET, INC. CONSTRAINED TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE EF 12 MM THICKNESS; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: mdr264920.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported from the wir form that persona knee instruments were returned and reported fractured and missing ball bearings.The instruments were noticed when they were returned to the office and were being checked in.There was no patient involvement.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The returned product exhibits signs of repeated use (nicked or gouged) and has one of components 1 and 2 disassembled / missing.The missing component was not returned.The device history record(s) were reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTRAINED TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE EF 12 MM THICKNESS
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11108811
MDR Text Key225299883
Report Number0001822565-2021-00003
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
K123459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42527900402
Device Lot Number62627134
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/17/2020
Initial Date FDA Received01/04/2021
Supplement Dates Manufacturer Received03/08/2021
Supplement Dates FDA Received03/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-