• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL TACTICATH CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED BID CURVE D-F; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL TACTICATH CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED BID CURVE D-F; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number A-TCSE-DF
Device Problem Entrapment of Device (1212)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/30/2020
Event Type  malfunction  
Event Description
During a vt ablation with tacticath se, the catheter was advanced via a retrograde aortic approach into the left ventricle for ablation on the posterior wall.During positioning of the catheter, the catheter advanced through the mitral valve and subsequently became entangled with the chordae tendonae.While trying to pull the catheter back, high force was noted on the tacticath se.As a result, a structual physician was called to consult on trying to remove the catheter and introduced a multipurpose catheter to loosen the hold of the chordae.During this, further manipulation was attempted on the tacticath se by the electrophysiologist and the catheter was freed without consequence to the patient.No further ablation was done and post-testing showed a successful procedure with no adverse consequences due to the entaglement of the catheter.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported entanglement and subsequent intervention remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED BID CURVE D-F
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
MDR Report Key11113396
MDR Text Key225565683
Report Number3005334138-2021-00002
Device Sequence Number1
Product Code OAE
UDI-Device Identifier05415067027641
UDI-Public05415067027641
Combination Product (y/n)N
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Model NumberA-TCSE-DF
Device Catalogue NumberA-TCSE-DF
Device Lot Number7718614
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/30/2020
Initial Date FDA Received01/04/2021
Supplement Dates Manufacturer Received01/25/2021
Supplement Dates FDA Received01/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight81
-
-