Model Number TABLETOP
Device Problem Computer Operating System Problem (2898)
Patient Problem Insufficient Information (4580)
Event Date 12/11/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that before surgery, the system froze.The system was switched out to initiate the case.
 
Manufacturer Narrative
The system was examined and the reported event was replicated.Two hard disk (hd) drives were subsequently replaced to resolve the issue.The system was tested and found to meet product specifications.The manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to nonconforming hard disk drives.Data will continue to be monitored for evidence of adverse trending with further action taken, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key11113611
MDR Text Key226065779
Report Number2028159-2021-00007
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/13/2020
Initial Date FDA Received01/04/2021
Supplement Dates Manufacturer Received04/14/2021
Supplement Dates FDA Received04/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-