This is filed to report the gripper actuation issue.It was reported that during preparation of the mitraclip delivery system (cds) the gripper arms did not fully lower, they stayed at approximately 60 degrees.The clip also failed to establish final arm angle (efaa).The clip opened during efaa to 30-40 degrees.There was no patient involvement, the cds was not used and was replaced.There was no clinically significant delay during the procedure.No additional information was provided.
|
The device was returned for analysis.The returned device analysis did not confirm the reported difficult to open or close/gripper actuation and unintended movement (efaa) as the device functioned as intended.Potential causes for gripper actuation issue are due to unintended curves on the device applied by the user or damage to the internal components caused by increased tension on the system and actuating the grippers more often than needed or not following the steps outlined in the instruction for use (ifu).Potential causes for the reported unintended movement include, but are not limited to, extreme curves applied to the system, failing to re-lock the clip by advancing the lock lever or not following the steps outlined in the instructions for use.The discrepancy between what was reported vs what was noted during the return device analysis is likely due to user technique/procedural circumstances possibly resulting from the curves/tension on the device during preparation a review of the lot history record revealed no manufacturing nonconformities to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from the reported lot.All available information was investigated and a conclusive cause for the reported difficult to open or close and unintended movement could not be determined in this event.There is no indication of a product issue with respect to manufacture, design, or labeling.
|