• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL STEM Back to Search Results
Catalog Number UNK HIP FEMORAL STEM
Device Problems Use of Device Problem (1670); Osseointegration Problem (3003)
Patient Problems Ossification (1428); Foreign Body Reaction (1868); Bone Fracture(s) (1870); Pain (1994); Inadequate Osseointegration (2646); Joint Laxity (4526); Muscle/Tendon Damage (4532); Physical Asymmetry (4573); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 04/03/2019
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Clinical symptoms code: appropriate term/code not available (b)(4) used to capture infections.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Literature article entitled, ¿zirconia phase transformation in zirconia-toughened alumina ceramic femoral heads: an implant retrieval analysis¿ by thu m.Nguyen, ms, et al, published by the journal of arthroplasty, volume 34, issue 12, december 2019, pages 3094-3098, was reviewed.The study examined the effects of in vivo and artificially-aged phase changes on the burst strength of zirconia-toughened alumina ceramic femoral heads.Twenty-eight zirconium composite alumina ceramic heads were retrieved during revisions, involving four different manufacturers, three different diameters (3 were 28 mm in diameter, 9 were 32 mm, and 16 were 36 mm).It was also noted that other products also revised, affecting four different manufacturers: regarding depuy, there was 1 marathon xlpe and 1 altrx xlpe and 1 non-xlpe polyethylene liner, and 1 tri-lock stem and 1 srom stem.This suggests that there were possibly five ceramic femoral heads from depuy revised, of the twenty-eight total.The following reasons for revision and the number of associated femoral heads retrieved was provided by the authors: loosening: 8, altr: 3, infection: 3, dislocation: 1, fracture: 4, malpositioning: 1, abductor tear: 1, heterotopic ossification: 1, leg length discrepancy: 1, pain: 1, unknown: 4.No other details were provided, nor was information provided that associated manufacturers with the various reasons for revision.Additionally, authors reported that an unspecified number of retrieved femoral heads had metallic scoring, indicative of impingement of the femoral head with the edge of the titanium acetabular cup.Results: although showing evidence of phase change, the heads achieved a burst load far above the 46 kn food and drug administration acceptance criterion, with 3 having burst strengths exceeding 100kn.The study showed that phase transformation occurs in vivo in ceramic composite femoral heads, but the amount transformed does not increase with the length of time the head are implanted.The effect upon burst strength of the retrieved heads is negligible.Impacted products: 1 unknown trilock stem, 1 unknown s-rom stem, 1 unknown s-rom augment, 1 unknown marathon xlpe liner, 1 unknown altrx liner, 1 unknown polyethylene liner, 5 unknown delta ceramic femoral heads.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: a device history record (mre) review, was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL STEM
Type of Device
HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11114513
MDR Text Key233124414
Report Number1818910-2021-00152
Device Sequence Number1
Product Code MRA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,literatur
Type of Report Initial,Followup
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL STEM
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/18/2020
Initial Date FDA Received01/04/2021
Supplement Dates Manufacturer Received02/02/2021
Supplement Dates FDA Received02/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-