This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The following sections were updated: d4, g4, g7, h2, h4, h6 and h10.The device was manufactured sometime in may 2013.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 7 years since the subject device was manufactured.Based on the results of the investigation, the physical damage observed on the device was most likely due to user mishandling.The instructions for use state, "excessive squeezing force on handles can lead to failure of forceps jaws.Check proper alignment of jaws/blades and telescope.Both should be visible when viewing through the telescope.Jaws/blades should be visible whether they are in the open or closed position.".
|