Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/18/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2021-00039.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported via worn instrument return form that the trial articular surfaces were returned fractured.The products were discovered during sterile processing.There was no patient involvement.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The returned item exhibits signs of repeated use and the anterior of the post feature has fractured off.All pieces were returned.The dhr was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.A definitive root cause cannot be determined.Evaluation of the returned device identified the fracture was consistent with the tasp fractures analyzed in a zrm.The zrm identified that the common failure modes for the tasp devices include either bending overload or low cycle fatigue culminating in bending overload as evident by the presence of hackle marks, river lines and striations features on the fracture surface.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|