• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 DXTEND STAND PE CUP D42 +3MM; DELTA XTEND IMPLANTS : SHOULDER HUMERAL CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 DXTEND STAND PE CUP D42 +3MM; DELTA XTEND IMPLANTS : SHOULDER HUMERAL CUP Back to Search Results
Model Number 130742203
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 12/18/2020
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).Health effect - clinical code: appropriate term / code not available (e2402) used to capture the infection (e19).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The patient was revised to address infection of spinal hardware/shoulder implants.Surgeon removed all shoulder implants and out in an antibiotic spacer.Upon opening the joint he found traces of metaloasis, surgeon was unsure what caused it.The glenosphere was loose and pulled out with a ronguer.The metaglene baseplate and screws all came out with the glenosphere.The inferior locking screw was broken midway on the screw.No further patient information is available at this time.Doi: (b)(6) 2017, dor: (b)(6) 2020, right shoulder.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed corrected: d3, g1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND STAND PE CUP D42 +3MM
Type of Device
DELTA XTEND IMPLANTS : SHOULDER HUMERAL CUP
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key11117471
MDR Text Key225091328
Report Number1818910-2021-00221
Device Sequence Number1
Product Code KWS
UDI-Device Identifier10603295027539
UDI-Public10603295027539
Combination Product (y/n)N
PMA/PMN Number
K120174
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number130742203
Device Catalogue Number130742203
Device Lot Number5281701
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/18/2020
Initial Date FDA Received01/05/2021
Supplement Dates Manufacturer Received02/05/2021
Supplement Dates FDA Received02/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DXTEND GLENOSPHERE STD D42MM.; DXTEND METAGLENE.; DXTEND MOD CENT EPI 1 HA.; DXTEND MODULAR HUM STEM D12 HA.; DXTEND SCREW LOCK D4.5X30MM.; DXTEND SCREW LOCK D4.5X30MM.; DXTEND SCREW NO LOCK D4.5X18MM.; DXTEND SCREW NO LOCK D4.5X18MM.; DXTEND GLENOSPHERE STD D42MM; DXTEND METAGLENE; DXTEND MOD CENT EPI 1 HA; DXTEND MODULAR HUM STEM D12 HA; DXTEND SCREW LOCK D4.5X30MM; DXTEND SCREW LOCK D4.5X30MM; DXTEND SCREW NO LOCK D4.5X18MM; DXTEND SCREW NO LOCK D4.5X18MM
Patient Outcome(s) Required Intervention;
Patient Age65 YR
Patient Weight121
-
-