• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EUROPEAN VALVE SET, EM2400; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EUROPEAN VALVE SET, EM2400; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938724E
Device Problem Packaging Problem (3007)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the packaging (overpouch) of a em2400 valve set was opened.This was identified before use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4, and h6.H4: device manufactured between may 12, 2019 to may 14, 2019.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.  visual inspection was performed to the photograph using the naked eye which revealed that the entire right side seal of the primary packaging was opened.The reported condition was verified. by the nature of the sample, no additional tests were performed. the cause of the condition could not be determined. a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EUROPEAN VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11117780
MDR Text Key225123685
Report Number1416980-2020-08309
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412475776
UDI-Public(01)00085412475776
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/31/2022
Device Catalogue NumberH938724E
Device Lot Number60185854
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/11/2020
Initial Date FDA Received01/05/2021
Supplement Dates Manufacturer Received02/08/2021
Supplement Dates FDA Received02/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-